command. The bottom panel shows information about the Pulsar 710.
4.4.2.4 Pulsar 710 Utility Dialog Box Output Controls:
4.4.2.4.1 Select Channel: The buttons labeled 1, 2, 3, and 4 select which
channelthesettingsaretoapplyto.SynchronizeandEqualizeoptionscan
be selected by checking the appropriate box.
•Synchronize:Whenthisboxischecked,anychangeinthesettingsfor
the selected channel are added to the three unselected channels.
•Equalize:Whenthisboxischeckedthesettingsoftheselectedchannel
are applied directly to the three unselected channels.
4.4.2.4.2 Program Button: None of the settings described below
will take effect until the Pulsar 710 is programmed. Pressing the program
button will transfer the settings from the host computer to the Pulsar 710
allowing the settings to take effect. The program button will be highlighted
to indicate that there are new settings that have not yet been transferred
tothePulsar710.IfthePulsar710isashinganerrorcodeuponinitial
start up a USB time out error will be experienced when you press the
program button on the software interface. If this situation occurs, turn off
the power switch on the Pulsar 710 and program. Once the error code
isclearedthePulsar710canbeprogrammed“onthey.”
4.4.2.4.3 Mode:Oneofvemodescanbeselectedhere.
• Pulsed: The Pulsar 710 produces pulses in response to a trigger signal.
The current and timing are determined by the control settings.
• Constant: The Pulsar 710 produces a constant output. The current is
determined by the control settings. The timing settings are not available
in this mode.
• Timing Bypass: The Pulsar 710 produces pulses that follow the trigger
signal. As an example, if the trigger pulse width is set to 10u secs
the output of the Pulsar 710 will turn the light on for 10u sec. It
is important to remember that Signatech II parameters always take
precedence over input pulse width. If the selected trigger is set to
falling edge the output is inverted. The current is determined by
the control settings: The timing settings are not available in this mode.
• Gated DC: The Pulsar 710 produces a constant output that is “Gated”
by the trigger signal. If the selected trigger is set to falling edge the
output is enabled by a low signal, logic “0” on the trigger, otherwise
a high signal, logic “1” enables the output. The timing settings are not
available in this mode.